Users Manual Part 1
Avoid exposure of your Pump to temperatures
below 41 ˚F (5 ˚C) or above 104˚F (40 ˚C).
Remodulin solutions freeze near 0 °C (32 °F)
and degrades at high temperatures. If you are
outside in cold weather, wear your pump close
to your body and cover it with warm clothing. If
you are in a warm environment, take measures
to keep your Pump and Remodulin cool.
Location of Use Warnings
Keep the system components, including pump
batteries, away from small children. Failure to
do so could result in children swallowing them
which can lead to choking or damage to their
digestive tract resulting in harm.
Portable and mobile RF communications
equipment (such as Walkie-Talkies, hand
held/vehicle mounted HAM, radios in Police
cars, Ambulances, and Firetrucks) may affect
the communication between the Pump and
Remote or interrupt operation of the Pump. If
the Remote indicates a loss of communication,
move the Pump and Remote at least 8.9m(30
ft) away from these items.
Do not sleep with the tubing set exposed if
you have pets.Doing so can result in damage
to the infusion set leading to interruption of
therapy and unintended exposure to
medication resulting in harm.
DKPI-09153-001-Rev. 1.4 23