Datasheet

Specifications subject to change without notice
ATM 160T Medical 150 W 160 W
Features
ANSI/AAMI/IEC/EN ES60601-1 (Edition 3.1)
EMC: 60601-1-2 (Edition 4.0)
IEC / EN 60950-1 Approval
Means Of Protection: 2x MOPP
Touch Current: < 100 µA
Regulated Output With Low Ripple Noise
1 Year Warranty
Marks / Approvals
Specifications
OUTPUT
MODEL
Voltage
Min. Load
Max. Load
Max. Power
Connector (mm)
ATM 160T-P120
12 V
0.00 A
12.5 A
150 W
4-pin Mini-DIN
ATM 160T-P190
19 V
0.00 A
8.4 A
160 W
4-pin Mini-DIN
ATM 160T-P240
24 V
0.00 A
6.6 A
160 W
4-pin Mini-DIN
ATM 160T-P480
48 V
0.00 A
3.3 A
160 W
4-pin Mini-DIN
INPUT
Voltage
AC 100-240V
Frequency
50-60Hz
Current (rms max)
1.8A-0.9A
Protection
Internal Primary Current Fuse
PROTECTION
Over Current
Yes
Overvoltage
Yes
Short Circuit
Yes
Over Temperature
Yes
Touch Current
< 100 µA
OTHERS
Topology
Switching Flyback
Dielectric Withstand
4000VAC Primary Secondary
Efficiency
DoE Level VI; Energy Star; ErP Stage 2 Certified
Ripple
2% Vp-p Max. for Output Voltage @ Full Load
Transient Response
0.5 ms for 50% Load Change Typical
Hold-Up Time
10 ms @ Full Load
Load Regulation
±5% Typical
Environment
Temperature
Operating: 0°C 40 °C (Storage: -20°C 80°C)
Humidity
Operating: 20 80 % RH (Storage: 10 90 % RH)
MTBF
SAFETY
EMI
EMS
Case Dimensions

Summary of content (2 pages)