Energy – Manual Mode (DDP-2001 Adult and DDP-2003 Non-attenuating Pediatric Defibrillation Pads) (Nominal)
Load Impedance (Ohms)
Selected
Energy
25 50 75 100 125 150 175
25 J 25 25 26 26 26 25 25
50 J 50 50 51 51 51 50 49
70 J 70 70 71 71 72 70 68
100 J 99 100 101 101 102 100 97
150 J 153 151 152 151 153 146 142
200 J 194 195 198 197 201 195 189
DDP-2003 pediatric defibrillation pads Selected Energy range is limited from 25 to 100 J.
Phase Time and Energy (DDP-2002 Attenuating Pediatric Defibrillation Pads)
Patient Impedance
(Ohms)
Phase A Duration
(ms)
Phase B Duration
(ms)
Energy Delivered
(Joules)
25 4.1 4.1 35
50 5.8 3.8 47
75 5.8 3.8 51
100 7. 2 4.8 53
125 7. 2 4.8 52
150 9 6 53
175 9 6 51
Applies to both AED and Manual Modes.
Note: If impedance is out of range for proper analysis and shock delivery, voice and/or visual prompts will inform the user.