6. ECG Monitor Mode
(DDU-2400/2450 only)
6.3 Monitoring The Patient
After the ECG monitoring electrodes are connected, the defibrillator displays a non-diagnostic ECG of the patient’s
heart rhythm and heart rate in a Lead II configuration. While in ECG Monitor Mode, the defibrillators shock capability
is disabled; however, the defibrillator continues to evaluate the patient’s ECG for a potentially shockable rhythm.
Monitoring
Monitoring
Monitoring
Plug in Rescue Pads
ECG Monitor Mode Screen Shockable rhythm detected
If a shockable rhythm is detected, the defibrillator prompts Plug in Rescue Pads.
• Confirm the patient’s condition: Not responsive? Not breathing? No signs of circulation?
• Remove the ECG Monitoring Adapter and connect the defibrillation pads (Rescue Pads) to the AED.
Apply the defibrillation pads to the patient’s chest, keeping them at least 2.5 cm (one inch) away from the
ECG monitoring electrodes. If necessary, remove the ECG monitoring electrodes.
• Follow the AED’s voice and onscreen text prompts.