7.2 Downloading the Internal Data Log
Regardless of whether a DDC is installed in the unit, select information is recorded internally within
the DDU-120 AED. The information recorded is limited to:
The time the AED was turned on.
Other data such as event milestones (motion detection, shock advice, shock delivery
information, etc).
Eight seconds of ECG data immediately before a shock/no-shock decision, eight seconds
immediately after each shock, and all ECG data during the charging and waiting-to-shock
periods.
Note: Audio data is not logged internally.
To download the internally logged information, perform the following procedure:
Insert a blank DDC into the unit.
Turn the unit on.
Once the unit is on, turn it off in data download mode by pushing and holding the ON/OFF
button for at least five seconds.
Allow the unit to write the contents of the internal log to the DDC by waiting for the unit to
turn off automatically.
The DDU-120 will write the contents of the internal log onto the DDC. This information can then
be reviewed using the DefibView software.