8.1.3 Defibrillator
Category Specification
Waveform
Biphasic Truncated Exponential
Energy
Adult: 150 Joules*
Infant / child: 50 Joules*
Charge control
Automatic by Patient Analysis System
Charge time from shock-advised
Typically < 6 seconds with a fresh DBP-2800 battery pack
and < 9 seconds with a fresh DBP-1400 battery pack.
Charge time may increase at the end of battery life and for
temperatures below 10°C.
Charge complete indication
SHOCK button flashing
“Press Shock button” voice prompt
Shock delivery
Shock is delivered by a single SHOCK button
DISARM
Automatic
If Patient Analysis System decides rhythm is no longer
shockable, or
Within 30 seconds after Charge complete, if operator
has not pressed SHOCK button, or
If defibrillation pads are removed from patient or
unplugged from unit.
Manual
If operator presses the OFF/DISARM button at any time
to disarm and turns off the device.
* nominal (±15%) delivered into a 50 ohm load.
8.1.4 Waveform Specifications
The DDU-100 AED delivers a 150J Biphasic Truncated Exponential waveform to patients with
impedances ranging from 25 to 180 ohms.
V
V max
V min
A
B
V1
V1'
t
The waveform is adjusted to compensate for measured patient impedance. Nominal phase times
and energy delivered are shown in the tables that follow.