Product Manual

10.2 Battery Packs
Use only Defibtech battery packs in the DDU-2000 Series AED.
DBP-2003 and DBP-2013 Battery Packs
Category Specification
Model number
DBP-2003
DBP-2013 (Aviation;TSO C-142a)
Main battery type
12VDC, 2800 mAh, Lithium/Manganese Dioxide.
Disposable, recyclable, non-rechargeable.
Capacity 125 shocks or 8 hours of continuous operation.*
Charge time 4 seconds or less (from shock advised)*
Standby-life (installed in unit) 4 years*
*Typical, new battery, at 25°C
Note: For information about optional rechargeable lithium-ion battery packs, please visit www.defibtech.com or
contact Defibtech or your distributor.
10.3 Self-Adhesive Defibrillation Pads
Defibtech self-adhesive defibrillation pads have the following characteristics:
Model number DDP-2001 DDP-2002
Type Adult Child/Infant < 8 years
< 55 lbs. (25 kg)
Intended use Disposable Disposable
Adhesion Self-adhesive Self-adhesive
Active gel surface area 12 inches
2
(77 cm
2
) each (nominal) 7.75 inches
2
(50 cm
2
) each (nominal)
Cable/connector type Integrated Integrated
Cable length 4 feet (122 cm) (typical) 4 feet (122 cm) (typical)
Expiration date 2.5 years from date of manufacture 2.5 years from date of manufacture
10.4 ECG Monitoring Adapter / Cable (optional)
For use only with DDU-2450 AEDs.
Model number DAC-2020 and DAC-2021
Patient connection Type CF, fully defibrillation-protected
Cable length 6.5 feet (2 meters)
Patient lead wire designation
Lead type DAC-2020 (AAMI) DAC-2021 (IEC) Placement
Positive Red - LL Green - F Left leg
Negative White - RA Red - R Right arm
Reference Black - LA Yellow - L Left arm
Typical lead connection Lead II
Performance with
DDU-2450 AED
Meets environmental specifications for DDU-2000 Series.
10. Technical Specifications
56
DAC-A2510EN-BG