Product Manual

Defibrillator – ECG Mode (DDU-2450 only)
Category Specification
Displayed ECG
ECG information is received from defibrillation pads in anterior-lateral or
anterior-posterior positions, or from the 3-wire ECG Monitor Adapter in Lead II
Screen type TFT Color LCD with backlight (53.6mm x 71.5mm) (320 x 240 pixels)
Displayed Range Differential: +/- 2 mV full-scale
Sweep Speed 25 mm/sec
Frequency Response 1 Hz to 22 Hz (-3 dB), nominal
Sensitivity 10 mm/mV, nominal
Heart rate display
20 to 300 bpm, updated once/second*
• Displayed when ECG viewing is enabled
• An out of range heart rate is indicated by dashed lines in the display
*On DDU-2000 Series units running software version 2.7 and higher.
Note: The ECG Display provides non-diagnostic ECG of the patient’s heart rhythm. It is not intended to provide
diagnostic or ST segment interpretation.
Note: The ECG waveform sensitivity and magnification settings are fixed and are not user adjustable.
Defibtech DDU-2000 Series AED (continued)
10. Technical Specifications
52
DAC-A2510EN-BG