10. Technical Specifications
10 Technical Specifications
10.1 Defibtech DDU-2000 Series AED
General
Category Specification
Size 7.3 x 9.5 x 2.3 inches (18.5 x 24 x 5.8 cm)
Weight Less than 3 lbs (1.4 kg) (with battery)
Power Battery Pack (not rechargeable)
Design standards
Meets applicable requirements of
• IEC 60601-1
• UL 60601-1
• CAN/CSA C22.2 No.601.1-M90
• IEC 60601-1-2
• IEC 60601-2-4
• AAMI DF80
Device Classification Internally powered with defibrillator-proof BF-type patient applied parts (per EN 60601-1)
Patient safety All patient connections are electrically isolated
Rescue protocol AHA/ERC (default); supports protocol updates by the user (password protected)
Defibrillator – AED Mode
Category Specification
Waveform Impedance Compensated Biphasic Truncated Exponential
Energy
Adult: 150 Joules (nominal [+/-15%] delivered into a 50 ohm load)
Child/Infant: 50 Joules (nominal [+/-15%] delivered into a 50 ohm load)
Charge control Automatic by Patient Analysis System
Charge time
4 seconds or less (from shock advised)*
Charge time may increase at the end of battery life and for temperatures
below 10°C.
Charge time from the
initiation of rhythm analysis to
readiness for discharge
Meets or exceeds AAMI DF80 and IEC 60601-2-4 requirements
Charge time measured from
initially switching power on to
charge ready
Meets or exceeds AAMI DF80 and IEC 60601-2-4 requirements
Charge complete indication
• SHOCK button flashing
“Press Flashing Shock Button” voice prompt and “Press SHOCK Button
text prompt
Shock delivery Shock is delivered by a single SHOCK button
DISARM
• If Patient Analysis System decides rhythm is no longer shockable
Within 30 seconds after charge complete, if operator has not pressed
SHOCK button
• If defibrillation pads are removed from patient or unplugged from unit
If operator presses and holds the ON/OFF button for approximately
two seconds, device will disarm and turn off
*Typical, new battery, at 25°C