Product Manual

7.8 Rescue Options Screen
To manually configure rescue options such as Rescue protocol and CPR breathing, select the Rescue options
from the AED Main Menu screen.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to Rescue options:
AED Main Menu Ú Rescue options
What it does: The unit will display the Rescue options menu screen. This screen will allow the user to modify
certain user changeable parameters:
CPR breathing
Rescue protocol
– Settings
Default view
To exit: Using the TOP or BOTTOM softkey buttons, scroll to and highlight the selection Go to main menu.
Press the CENTER softkey button. The unit will exit the Rescue options menu screen and return to the
AED Main Menu.
CPR Breathing
The CPR breathing option allows the user to set the AED's default CPR setting to include or not
include CPR breathing coaching prompts during CPR.
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to CPR breathing:
AED Main Menu Ú Rescue options Ú CPR breathing
What it does: CPR breathing allows the user to set the AED's default CPR setting to include or not
include CPR breathing coaching prompts.
Use the TOP and BOTTOM softkey buttons to select the desired mode. When the desired selection is
chosen, press the CENTER softkey button to set the feature. The AED will now use that coaching setting.
Note: The factory default setting of CPR breathing is set to Disabled.”
Rescue Protocol
The AED supports two rescue protocols at one time. The Rescue protocol option allows the user to
select a rescue protocol. Rescue protocol options include the AHA/ERC protocol or “Custom”
Before you begin: Be sure the unit is in Maintenance Mode.
To enter: Navigate to Rescue protocol:
AED Main Menu Ú Rescue options Ú Rescue protocol
What it does: The Rescue protocol option allows the user to select between up to two rescue protocols
that have been enabled in the unit. The factory default setting of the rescue protocol is “AHA.”
To change the protocol, press the CENTER softkey button to highlight the protocol. The user will be
prompted to enter a password to proceed. The password may be obtained from your medical director
or from Defibtech. (For Defibtech contact information, please refer to the “Contacts” section in
Chapter 13.) Once the password has been entered, the user may select between the two protocols.
To enter the password use the TOP softkey button to scroll through numbers. Once the correct number
appears, use the CENTER softkey button to move to the next space. Once all of the numbers have been
entered press the CENTER softkey button. The user will now be able to choose a different rescue protocol.
7. Maintenance Mode
45
DAC-A2510EN-BG