Item Brochure

Lifeline AUTO Fully Automatic Defibrillator
technical specifications*
Defibrillation/Monitoring Pads
MODEL
Adult – DDP-100
Child/Infant – DDP-200P
TYPE
Pre-connected, single-use,
non-polarized, disposable,
self-adhesive electrodes
with cable and connector
SURFACE AREA
103 cm
2
(nominal, each pad)
50 cm
2
(nominal, each pad)
PAD PLACEMENT
Adult – Anterior/Anterior
Child/Infant – Anterior/Posterior
CABLE LENGTH
(typical)
48 in (122 cm)
Event Documentation
INTERNAL EVENT RECORD
Critical ECG segments and rescue
event parameters are recorded
and can be downloaded to a
removable data card
PC-BASED EVENT REVIEW
ECG with event tag display,
and audio playback when available
REMOVABLE STORAGE
(optional)
Up to 12 hours of ECG and event
data storage (no audio option) or
up to 2 hours of audio, ECG and
event storage (audio option) on a
removable data card. Actual length
of storage is dependent on card
capacity
Environmental
TEMPERATURE
Operating:
0 to 50°C (32 to 122°F)
Standby (with battery installed):
0 to 50°C (32 to 122°F)
RELATIVE HUMIDITY
Operating / Standby: 5% – 95%
(non-condensing)
ALTITUDE
-500 to 15,000 ft (-150 to 4500 m)
per MIL-STD-810F 500.4 Procedure II
VIBRATION
Ground (MIL-STD-810F 514.5
Category 20)
Helicopter (RTCA/DO-160D,
Section 8.8.2, Cat R, Zone 2,
Curve G)
Jet Aircraft (RTCA/DO-160D,
Section 8, Cat H, Zone 2,
Curves B & R)
SHOCK/DROP ABUSE
TOLERANCE
MIL-STD-810F 516.5 Procedure IV
(1 meter, any edge, corner,
or surface, in standby mode)
SEALING/WATER
RESISTANCE
IEC60529 class IP54;
Splash Proof, Dust Protected
(Battery Pack installed)
ESD
EN 61000-4-2:2009 (open air up to
15kV or direct contact up to 8kV)
EMC (Emission)
EN 60601-1-2:2001 +A1:2006, limits
EN 55011:1998 +A1:1999 +A2:2002
Group 1 Level B, method
EMC (Immunity)
EN 60601-2-4:2003, limits
EN 61000-4-3:2002
Level 3 (10V/m), method
Patient Analysis System
PATIENT ANALYSIS
Automatically evaluates patient
impedance for proper pad contact.
Monitors signal quality and
analyzes patient ECG for shockable/
non-shockable rhythms
SENSITIVITY/SPECIFICITY
Meets AAMI-DF-39 specifications
and AHA recommendations
Physical
SIZE
8.5 x 11.8 x 2.7 inches
(22 x 30 x 7 cm)
WEIGHT
(Approximate)
With DBP-1400: 4.2 lbs (1.9 kg)
With DBP-2800: 4.4 lbs (2.0 kg)
Self Tests
AUTOMATIC
Automatic daily, weekly, monthly
and quarterly circuitry tests
BATTERY INSERTION
System integrity test
on battery insertion
PAD PRESENCE
Pads preconnected tested daily
USER-INITIATED
Unit and battery pack system test
may also be initiated by the user
STATUS INDICATION
Visual and audible indication
of unit status
MODEL DBP-2800 MODEL DBP-1400
POWER
15V, 2800 mAh
POWER
15V, 1400 mAh
CAPACITY
300 shocks or
16 hours continuous operation**
CAPACITY
125 shocks or
8 hours continuous operation**
STANDBY LIFE
7 years**
STANDBY LIFE
5 years**
Defibrillator
TYPE
Fully Automatic external defibrillator
MODEL
DDU-120A, DDU-120E
WAVEFORM
Biphasic Truncated Exponential
(Impedance compensated)
ENERGY
Adult: 150 Joules
Pediatric: 50 Joules
(Nominal into 50 Ohm load)
CHARGE TIME
**
DBP-2800: Less than 6 seconds
DBP-1400: Less than 9 seconds
VOICE PROMPTS
Extensive voice prompts
guide user through operation
of the unit and CPR
CPR PACING
Metronome
CONTROLS
Lighted On/Off button
INDICATORS
· “check pads”
· “do not touch patient”
· “analyzing”
· armed-for-shock
· AED status LED
**Typical, with new battery at 25° C
Battery Pack
TYPE
Lithium/Manganese Dioxide
Disposable, recyclable,
non-rechargeable
LOW BATTERY INDICATORS
Visible
Audible
**Typical, with new battery at 25° C
*Specifications subject to change without notice
Debtech, LLC • Guilford, CT 06437 1-203-453-4507
1-866-DEFIB-4U (866-333-4248) • www.debtech.com
0197
US
DAC-702-EN-BD
ELECTRONIC DISTRIBUTION