Instructions

Features
n Simple, easy-to-use interface
n 6 K-type thermocouple ports (channels)
n Measures and records up to 40,000 datasets (240,000 readings) in
up to 6 batches
n Records measurements every 2 seconds for up to 22 hours of
recording time
n Download, view, and analyze batch data using PosiSoft Desktop software.
Export .CSV files for easy import into spreadsheets.
n
NEW Cure Index compares supplier’s cure specifications to actual oven
and part temperatures to ensure a proper cure
n Bright LEDs indicate logger status at a glance
n High temperature stainless steel barrier box with fiberglass insulation for
long run times at high temperatures. Contains no silicone.
n Certificate of Calibration showing traceability to NIST included
Oven Temperature Logger
The PosiTest OTL Oven Temperature Logger with included barrier
box and holder is ideal for measuring oven temperature profiles.
Size: 30.5 x 25.5 x 15 cm Weight:* 9 kg
(12 x 10 x 6) (20 lbs.)
Power Source:
2 AAA batteries
Battery Life:
30 hours (recording time)
* Weight includes the PosiTest OTL, Holder, and Barrier
Box. Holder XT increases weight to 11 kg (25 lbs.).
The PosiTest OTL is available in two configurations. The PosiTest
OTL XT features an enhanced holder with heatsink for greater
endurance at high temperatures.
Two Kits available
PosiTest OTL Kit
PosiTest OTL XT Kit
The chart and table above represent the combined thermal characteristics
of the PosiTest OTL, Barrier Box and choice of Holder/Holder XT
Temperature Probes
n Wide variety of temperature probes available with 1.8, 3.6, or 6 m
(6’, 12’, or 20’) stainless steel braided cables (sold separately). See the
following page for details.
n Certificate of Calibration showing traceability to NIST included
Ordering Guide
PosiTest OTL Kit
OTLKIT
PosiTest OTL XT Kit
OTLKITXT
LOGGER KITS COME with temperature logger, stainless steel
high temperature barrier box, choice of holder, 6 engraved probe
ID tags, two AAA batteries, instructions, Long Form Certificate of
Calibration traceable to NIST, carrying case, USB cable, PosiSoft
Desktop software, and two year warranty.
PosiTest OTL Specifications
Channels
6 K-type non-grounded
thermocouples
Thermocouple Measurement Range
-100 to 1000° C
-148 to 1832° F
Logger Operating
Temperature Range
PosiTest
OTL without Holder or Barrier Box
0 to 70° C
32 to 158° F
Logger Accuracy
+
0.5° C
+
0.9° F
Resolution
0.1° C
0.1° F
Memory
40,000 Datasets. Each Dataset
consists of up to 6 channels.
Sample Interval
2 seconds (Fixed)
Maximum recording time is 22 hours
PosiTest OTL XT
(OTLKITXT)
300 minutes
200 minutes
140 minutes
100 minutes
150
° C
(302
° F)
200
° C
(392
° F)
250
° C
(482
° F)
300
° C
(572
° F)
PosiTest OTL
(OTLKIT)
90 minutes
65 minutes
50 minutes
not recommended