Owner's manual
30
Technical Data
Conforms to the practices of: BS3900(C5), ISO
2178/2360/2808, ASTM B499/ D1186/D1400,
SSPC-PA2
Range:
Microprobes, NAS
0-25 mils & 0-625 microns
FT, FTS, NT,
NTS, FNT, FNTS
0-250 mils & 0-6 mm
FHS, NHS
0-750 mils & 0-20 mm
All others
0-60 mils & 0-1500 microns:
Default Resolution:
Microprobes, NAS
0.01 mil / 0.5 µm
0.1 mil / 2 µm (> 4 mils / 100 µm)
FT, FTS, NT,
NTS, FNT, FNTS
0.5 mil / 0.01 mm
FHS, NHS
1 mil / 0.02 mm
All others
0.05 mil / 1 µm
0.1 mil / 2 µm (> 2 mils / 50 µm)
Accuracy:
Microprobes, NAS
±(0.01 mil + 1%) 0 - 4 mils
±(0.1 mil + 3%) > 4 mils
±(0.5 µm + 1%) 0 - 100 µm
±(2 µm + 3%) > 100 µm
FT, FTS, NT,
NTS, FNT, FNTS
±(0.5 mil + 1%) 0 - 100 mils
±(0.5 mil + 3%) > 100 mils
±(0.01 mm + 1%) 0 - 2.5 mm
±(0.01 mm + 3%) > 2.5 mm
FHS, NHS
±(1 mil + 1%) 0 - 250 mils
±(1 mil + 5%) > 250 mils
±(0.02 mm + 1%) 0 - 6 mm
±(0.02 mm + 5%) > 6 mm
All others
±(0.05 mil + 1%) 0 - 2 mils
±(0.1 mils + 1%) > 2 mils
±(1 µm + 1%) 0 - 50 µm
±(2 µm + 1%) > 50 µm