5.1.4 Save single values (Single)
A sequence of individual measuring values can be saved by pressing the S-button during the
measurement. Single values can be capture either during one or several measuring cycles. The
single value memory function always captures the actual displayed value with each press on
the S button as long as the measuring is active (dimensional unit flashes).
5.1.5 Save measuring cycles (Series)
Press the S-button to start the continuous recording of a measuring cycle, while the measuring
is active (N flashes). The recording is stopped either by pressing the S-button, the I-button
or automatically with the AUTO-STOP function. The entire sequence of the displayed values
(according the refresh rate) is captured and if the high speed mode is activated the absolute
peaks are saved also. The memory can register up to 1000 values.
5.1.6 Display statistical results
After finishing the (last) measurement, the results and statistcs can be displayed by pressing
the S-button.
S
S
S
S
S
S
S
Mem
cN
Mem
cN
Mem
cN
Mem
cN
Mem
cN
Mem
cN
Avg
Mem
cN
Dev
Last measured value
Max. measured value compression
Min. measured value compression
Max. measured value tension
Min. measured value tension
Average (arithmetic average)
Standard deviation
5.1.7 Delete results (memory)
In order to delete one of the saved values you press the 0-Button while the value is displayed.
The display now shows [0.000]. To delete all saved values at the same time press the 0-Button
for 2 seconds. The display returns to the initial screen [- . - - - ]
To start the next measurement press the I-button again.