Specifications

7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
B. Measuring the Amplitude Threshold
Execution of step 2 on the previous page displays the threshold test parame-
ter setup options as shown in this example.
1. Use the [
] and [
] buttons to adjust each of the parameter values to
the desired test setting.
Lower Rate
— Adjust to a value high enough to effect consistent
pacing. Loss of capture cannot be determined if the pacemaker is
inhibited.
Paced/Sensed AV
— If the test mode is DDD, adjust this interval to a
value short enough to provide a consistently paced ventricular rhythm.
The selected value applies to both Paced AV and Sensed AV intervals.
Pulse Width
Adjust to the pulse width value you want in effect
during the amplitude test. This value cannot be less than 1.0 ms.
Amplitude
— Adjust to the value at which you want the amplitude test
to start. This value must be high enough to provide consistent capture.
Pulses Per Decrement
— Adjust to the number of pacing cycles you
want to occur at each test value before reduction occurs (see page 7-4).
2. Position the programming head and hold it steady in position for the
remainder of this procedure. Observe that Marker telemetry appears
below the ECG trace.
3. Press and hold [
PROGRAM
] to execute the amplitude threshold test.
The threshold test execution window shows the temporary values in
effect. The arrow (<<<===) points to the value being reduced.
Observe the patient’s ECG trace as the programmer automatically
reduces amplitude one programmable step every 3 to 15 pacing cycles
(as selected in step 1). An audible “beep” indicates confirmation of
each new amplitude value.
4. When capture is lost, immediately release [
PROGRAM
].
7-6 Using the Programmer