Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
3. Repeat steps 1 and 2 to include any other parameters you want to
program.
NOTE: If you try to exit the screen before you program the selected
values, a pop-up window will provide the option to continue with this
procedure or clear the pending values and exit the screen.
4. Position the programming head and select [
PROGRAM].
Select the [
PROGRAM] button to program all pending parameter values.
Hold the programming head in position until the confirmation message
appears.
NOTE: If the programming is interrupted, it is possible that one or
more of the parameter groups will not be programmed and will still
show pending values. In such cases, the programming will not be
confirmed, and the message box will direct you to reselect [
PROGRAM].
Programming Mode Switch — If the present setting for Mode Switch is
ON or if you select ON, the value window lists the present setting for the
subordinate parameter Detect Rate (see example below) and provides the
option to select a new value. Refer to Chapter 14 for information about the
operation of Mode Switch.
1. To select a value for Detect
Rate, use the same basic
procedure described in steps
1 and 2 of the general
programming procedure.
2. When you have selected the
desired value for Detect
Rate, select the [
ACCEPT
VALUE] button to close the
primary parameter value
window.
If the present or pending value for Upper Tracking Rate is 165 ppm or
above, the programmer automatically selects a pending value for Detect
Rate. The values of 185 and 190 bpm are not selectable but will respective-
ly appear as the pending value if Upper Tracking rate is 175 or 180 ppm.
PROGRAM
Programming Pacing Parameters 6-5