Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Parameter Interaction — If you select a restricted parameter value, the
display will indicate the interacting parameter values by displaying the
values in reverse video. You must select new values to resolve the conflict
before programming is allowed.
Parameter Entry Requirement — A solid, shaded rectangle appearing
where a parameter value should be displayed indicates that a valid value for
this parameter must be selected before programming is allowed.
General Parameter Programming Procedure
For specific parameter information related to
each pacemaker model, refer to Sections B
and E of Part III.
To program one or more parameter values:
1. Select the desired parameter from those
displayed —
Lower Rate
for example.
Press the selector pen directly over the
parameter name or its value.
NOTE: Select
Pacing Mode
first if you
intend to program a mode. Selecting a
mode displays the programmable parame-
ters pertinent to that mode.
2. From the window of value options,
select the desired value.
Except for Hysteresis, selecting a value in
italics results in a restricted parameter
interaction.
Selecting a value closes the window. The
selected value appears in place of the
present value and is denoted as “pending”
by a dashed rectangle.
To clear a pending value from the screen,
select that parameter or value, and then
select [
CLEAR VALUE
] in the value win-
dow. To close the value window without
selecting a value, press [
ACCEPT VALUE
].
6-4 Using the Programmer