Specifications

7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Features of the Pacing Parameters Screen
Parameter Display — Following an interrogation, the screen shows the
present pacing parameter settings. The screen displays only those parame-
ters that are applicable to the selected pacemaker model and pertinent to the
present pacing mode. Note that the parameters are divided into labeled
groups according to their general function.
[
PROGRAM
] Button — This button appears in the lower right corner of the
screen only when one or more new parameter values have been properly
selected (see “Restricted Parameter Interactions” below). The
PROGRAM
key on the programming head will respond only when the [
PROGRAM
]
button shows on the screen.
Special Options — The three buttons located at the right of the parameter
display may be useful during the programming procedure described on the
next page.
When you select new parameter values to program, these
“pending” values are displayed in place of the interrogated
present settings. To temporarily recall the present settings,
press and hold this button.
If you have saved a set of parameter values (see below), this
button will appear on the screen. Pressing and holding this
button will temporarily display these saved values.
Selecting this button displays options for: 1) clearing all
pending values from the parameter screen, 2) saving a set of
parameter values, and 3) recalling the set of saved values or
nominal values to the screen as pending values. (Refer to
“Programming Nominal and Saved Values” in this chapter.)
Restricted Parameter Interactions
Certain combinations of parameter values are restricted because they are
invalid or result in undesirable interactions. The programmer recognizes
these combinations and does not allow programming until all parameter
conflicts are resolved and all parameter selection requirements are met.
Refer to Section E in Part III for more information on parameter program-
ming requirements and restrictions.
Restricted Parameter Values — Any parameter value that conflicts with
the setting of a present or pending parameter is displayed in italics in the
parameter value option window (see step 2 of the programming procedure).
SHOW
PRESENT
SHOW
SAVED
SAVE/
NOMINAL
OPTIONS...
Programming Pacing Parameters 6-3