Specifications

7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Programming Pacing Parameters
To program changes to the permanent pacing parameters, begin by
selecting the Pacing Parameters screen. This screen displays the present
parameter settings and lets you select the parameters and values to be
programmed.
Selecting the Pacing Parameters Screen
You can use either of two methods to select the Pacing Parameters screen:
select the [
PACING PARAMETERS
] button if it is displayed in the lower right
corner of the present screen, otherwise select the [
PARAMETERS
] Menu
button, and then select the
PACING PARAMETERS
option as shown below.
Present settings will not be shown unless an appropriate interrogation has
occurred. Interrogate the pacemaker, if required, to display these settings.
1. Select [
PARAMETERS
]
menu button.
2. Select
PACING
PARAMETERS
option.
Pacing Parameters
Screen
Refer to Part II of this
manual for specific
information about the
pacing modes and the
operation of certain
parameters.
Refer to Section E in
Part III for a list of the
parameters and values
available for each
6-2 Using the Programmer