196407001 Rev A
Displaying a Marker Channel
™
Diagram
If the selected pacemaker is a Prodigy DR, Prodigy D, or Prodigy VDD
model, the Freeze screen includes the [
SHOW DIAGRAM
] button as shown in
the example on page 5-10. Under the proper conditions (see NOTE below),
selecting this button displays a Marker Channel Diagram.
NOTE: A Marker Channel Diagram can be displayed only if the pacing
mode is dual chamber or VDD and the display includes Marker telemetry
annotations (the selected telemetry type can be either Marker or EGM).
Selecting the Marker Channel
™
Diagram
To display a Marker Channel Diagram, select the
[
SHOW DIAGRAM
] button on the Freeze screen.
The diagram will be drawn on the lower trace
display as shown in the example below.
To return to the trace of Marker or EGM
plus Marker telemetry, select the [
SHOW
MARKERS
/
EGM
] button, which appears in
place of the [
SHOW DIAGRAM
] button.
SHOW DIAGRAM
Freeze Screen with
Marker Channel™
Diagram
To print the ECG trace
and diagram, select the
[
PRINT
] menu button, and
then select the
CURRENT
SCREEN REPORT
option.
SHOW MARKERS/EGM