Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Programming the Trace Telemetry Type
Default Telemetry
—
Positioning the programming head over the patient’s
pacemaker automatically causes the pacemaker to transmit both Marker
telemetry and EGM telemetry (atrial EGM for dual chamber modes). The
combination of Marker telemetry and EGM telemetry is displayed in the
examples on pages 5-3 and 5-5.
Programming Procedure — Use the following procedure to temporarily
program the telemetry to another type:
1. With the programming head in position over
the pacemaker, select this button.
2. Select the desired telemetry type from the
options displayed (see below).
3. Continue to hold the programming head in
place for as long as you want to view the
programmed telemetry.
The telemetry type will return to its default
setting if the programming head is removed
from the pacemaker site for over 30 sec-
onds.
Telemetry Options
—
The programmable options for the type of telemetry
that can be displayed and recorded with the patient’s ECG are as follows:
A. EGM
This option displays atrial EGM and Marker telemetry.
V. EGM
This option displays ventricular EGM and Marker
telemetry.
MARKER
This option displays Marker telemetry only. The Marker
trace appears in a larger format on the lower trace display
in place of the EGM.
SENSOR
This option displays Sensor telemetry only. The Sensor
trace appears on the lower trace display in place of the
EGM.
ECG LEAD II
1.0 mV/mm
Filter ON
75 bpm
V. EGM
1.0 mV/mm
Displaying and Printing the Patient’s ECG and EGM 5-7