Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Printing the Patient’s ECG and EGM
At any time during a patient session, you can initiate a continuous, real-time
recording of the patient’s ECG as described below.
Procedure
The procedure is slightly different depending on whether you are operating
a 9760 or 9790 programmer.
NOTE: Step 1 will interrupt any report being printed. Likewise, selecting a
Print option from the display screen will interrupt and cancel an ECG
recording in progress.
9760 Programmer
9790 Programmer
1. To start a recording of the patient’s
ECG:
9760 — Press the printer
RECORD
button.
9790 — Press the desired paper
speed button.
2. To stop the recording, press the
same button again.
3. Before you tear off the ECG strip,
press the
PAPER ADVANCE
button to
advance the paper to a perforation.
To change the recording speed on a 9760 programmer, press the desired
PAPER SPEED
button. This can be done prior to or during the recording
process.
NOTE: Because the printed waveform recording provides a higher
resolution, it may show artifacts and events that do not appear on the
display screen.
5-2 Using the Programmer