Specifications

7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Select the [MORE INFO] button for information about estimated time to
replacement and the estimated maximum, average, and minimum values.
IMPORTANT: Elective pacemaker replacement should not be based on
the estimated time to replacement. For this decision, use only the elective
replacement indicators or the battery status “Replace Pacer” message. When
battery voltage indicates that it is time for pacemaker replacement, the
“Replace Pacer” message automatically appears on the screen in place of
the estimated time to replacement value.
The following messages can appear in place of the estimated time value:
Select MORE INFO ===>>>” — Estimated time to replacement in-
formation is available only by selecting the [
MORE INFO] button. This
message appears when additional factors that affect the time estimate are
present and require an explanation. If less than 24 hours have elapsed since
the previous patient session, a message will state that there is “insufficient
event summary data accumulated in the pacemaker.” In this case, the
estimated time to replacement calculations assume that the patient is paced
100 percent of the time.
Replace Pacer” — Low battery voltage in the patient’s pacemaker has
tripped the ERI (elective replacement indicator). Elective replacement of
the pacemaker is now recommended.
Retrieving Pacing System Data 4-11