Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Battery/Lead
This section of the screen includes a summary of pacemaker battery and
lead status.
Estimated Time to Replacement
“Estimated Time to Replacement” is a calculated estimate of the time (in
months) remaining until pacemaker replacement will be required. This
calculated estimate is an average value based on the programmed parameter
settings and the event summary data accumulated by the pacemaker since
the previous patient session.
Select the [
MORE INFO
] button for information about estimated time to
replacement and the estimated maximum, average, and minimum values.
Refer to page 4-11 for information about the messages that can appear in
place of the estimated time value.
IMPORTANT: Elective pacemaker replacement should not be based on
the estimated time to replacement. For this decision, use only the elective
replacement indicators or the battery status “Replace Pacer” message.
When battery voltage indicates that pacemaker replacement is required,
the “Replace Pacer” message automatically appears on the screen in place
of the estimated time to replacement value.
Lead Impedance
“Lead Impedance” is the measured electrical impedance presented by the
pacing lead and electrode/tissue interface. If the pacing mode is a dual
chamber mode, the measured impedance for both lead systems is displayed.
Diagnostic Parameters
This section of the screen indicates the programmed status of the Diagnos-
tics function by displaying the present settings of the Setup and Type
parameters. No value for Setup will be displayed if any changes were made
to the default settings established by the Setup option. A setting of “Off” for
the Type parameter indicates that the Diagnostics function has not been
programmed to collect data.
4-6 Using the Programmer