Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
A Note About Mode Switch — The subordinate parameter Detect Rate
associated with programming Mode Switch does not show on the Pacing
Parameters screen. To view the setting of this parameter, select Mode
Switch to open its value window. If Mode Switch is set to ON, the present
value for Detect Rate will show in the Mode Switch value window.
Displaying Diagnostic Parameter Settings
To display the present programmed parameter settings controlling the
Diagnostics function, select the [
PARAMETERS
] menu button, and then
select the
DIAGNOSTIC PARAMETERS
option as illustrated below.
The Diagnostic Parameters screen appears as shown in the example below.
If the Diagnostics function has not been programmed to collect data, the
Type parameter will show a value of OFF.
1. Select [
PARAMETERS
]
menu button.
2. Select
DIAGNOSTIC
PARAMETERS
option.
Diagnostic Parameters
Screen
Interrogate the pace-
maker, if required, to
display settings. Refer
to Chapter 8 for instruc-
tions on using the
Diagnostics function.
Retrieving Pacing System Data 4-3