Specifications

7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Displaying Present Parameter Settings
You can quickly display the parameter settings to which the patient’s
pacemaker is programmed by selecting the appropriate parameter screen.
There are three such screens: 1) The Pacing Parameters screen, 2) The
Diagnostic Parameters screen, and 3) The Additional Parameters screen.
Displaying Pacing Parameter Settings
To display the Pacing Parameters screen, select the [
PACING PARAMETERS
]
button if it is displayed in the lower right corner of the present screen,
otherwise select the [
PARAMETERS
] menu button as illustrated below.
Parameter settings will not show unless an appropriate interrogation has
occurred. Interrogate the pacemaker, if necessary, to display the settings.
1. Select [
PARAMETERS
]
menu button.
2. Select
PACING
PARAMETERS
option.
Pacing Parameters
Screen
If the pacemaker has
been interrogated, this
screen displays the
presently programmed
pacing parameter
settings.
To use this screen to
program new pacing
parameter settings,
refer to the instructions
4-2 Using the Programmer