Specifications

7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Event Modes Timing Description
20. PVC (refractory DDD, DDDR, Starts minimum PVARP of 400 ms; starts the
V-sense not DDI, DDIR, VDD associated programmed PVAB period,
preceded by programmed Ventricular Refractory (including
A-sense, A-pace the associated 50 to 100 ms ventricular
or A-sense in blanking) period, and Upper Tracking Rate
PVARP) (DDD, DDDR, and VDD only).
21. V-sense in SAV DDD, DDDR, Starts VA interval, Upper Tracking Rate,
VDD PVARP (including the associated PVAB
period), and programmed Ventricular
Refractory (including the associated
50 to 100 ms ventricular blanking) period.
22. V-sense in PAV DDD, DDDR, Starts VA interval, Upper Tracking Rate
outside 110 ms DDI, DDIR, DVI, (DDD and DDDR only), PVARP (including
VSP window or DVIR the associated programmed PVAB period,
VSP is OFF but not in DVI and DVIR), and programmed
Ventricular Refractory (including the
associated 50 to 100 ms ventricular blanking)
period.
23. V-sense in PAV DDD, DDDR, Starts Upper Tracking Rate (DDD and DDDR
inside 110 ms DDI, DDIR, DVI, only), PVARP (including the associated
VSP window with DVIR programmed PVAB period, but not in DVI and
VSP is ON DVIR), and programmed Ventricular
Refractory (including the associated 50 to
100 ms ventricular blanking) period. Emits
safety pace at 110 ms or at PAV timeout,
whichever is less. Starts VA interval at the
Safety Pace.
24. V-sense after DDIR, DDI Restarts VA interval, programmed Ventricular
A-sense and before Refractory (including the associated 50 to
V-pace occurs 100 ms ventricular blanking) period, and
PVARP (including the associated programmed
PVAB period).
25. V-sense in DDD, DDDR, Restarts Upper Tracking Rate (DDD, DDDR,
refractory period DDI, DDIR, VDD and VDD only), PVARP (including the
(not PVC) associated programmed PVAB period),
and programmed Ventricular Refractory
(including the associated 50 to 100 ms
ventricular blanking) period.
26. V-sense in DVI, DVIR, VVI, Restarts programmed Ventricular Refractory
refractory period VVIR, VVT (including the associated 50 to 100 ms
ventricular blanking) period and Upper
Activity Rate (VVIR only).
E-14 Prodigy Reference Information
9879EN-E_9501043EN/CTC.qxd 5/10/99 11:40 Page E-14