Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Single Chamber Pacemaker Permanent
Programmable Modes and Parameters
In the table below are the Prodigy SR and Prodigy S permanent programmable modes and parameters.
Parameters listed under separate columns apply only to the specific Prodigy model. Parameters
common to both Prodigy models are listed once.
Parameter Prodigy SR Prodigy S Notes
(8158)
(8160/61/62) (8164/65/66)
Modes VVIR, VVI, VVT, VOOR VVI, VVT, VOO
VOO, AAIR, AAI, AAT AAI, AAT, AOO,
AOOR, AOO, OVO, OAO OVO, OAO
Lower (Pacing) Rate 30, 35, 40, … 180 ppm (except 65 and 85 ppm)
Upper Activity Rate 80, 90, 95, 100, … 180 ppm Not Applicable Rate responsive modes
Refractory Period 150, 160, 170, … 500 ms
Amplitude* 0.5, 1.0, 1.5, 2.0, 2.5, 3.0, 3.5, 4.0, 5.0, 7.5 V
Pulse Width 0.03, 0.06, 0.09, 0.12 ms
0.15, 0.20, 0.25, … 1.0 ms
1.1, 1.2, 1.3, 1.4, 1.5 ms
Sensitivity 0.25 and 0.35 mV
Atrial 0.25, 0.35, 0.5, 0.7, 1.0, 1.4, 2.0, 2.8, 4.0 mV sensitivity for bipolar
Ventricular 1.0, 1.4, 2.0, 2.8, 4.0, 5.6, 8.0, 11.2 mV sensing only
Blanking 100, 110, 120, … 350 ms AAIR, AAI, and
AAT modes
Pacing Polarity Bipolar, Unipolar Models 8161 and 8165
unipolar pacing and
Sensing Polarity Bipolar, Unipolar sensing only
* Tolerance is based on 37°C and a 510 ohm load for the following: 0.5 V (± 10%), 1.0 V (± 10%), 1.5 V (± 10%),
2.0 V (± 10%), 2.5 V (+ 2/- 13%), 3.0 V (± 10%), 3.5 V (+ 9/- 7%), 4.0 V (± 10%), 5.0 V (+ 1/- 13%), 7.5 V (+ 1/- 13%).
E-6 Prodigy Reference Information
9879EN-E_9501043EN/CTC.qxd 5/10/99 11:40 Page E-6