Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Single Chamber Pacemaker Diagnostics
NOTE: Prodigy diagnostics are programmed using the diagnostic setups
listed below that set the diagnostic type and associated parameters listed
under “setup values”. Alternately, other programmable values can be
selected for the diagnostic type and parameters.
Diagnostic Setup Setup Values Other Programmable Values
Tachy Monitor
3
Type: High Rate Episode
Method: Rolling Frozen
Detection Rate
4
: Upper Activity Rate or 80 - 400 ppm
120 ppm, whichever is greater
Detection
Number: 3 2- 200
Termination
Number: 3 2 - 20
Collect EGM: ON OFF
Lead Monitor: OFF ON,
ON + Polar. SW.
Rate Histogram
Type: Rate Histogram
Refractory: Include Exclude
Lead Monitor: OFF ON,
ON + Polar. SW.
Lead Trend
Type: Lead Trend
Method: Rolling Frozen
Lead Monitor: OFF (fixed setting)
3
Diagnostic is only applicable to the Prodigy SR pacemaker.
4
The value used for the Upper Activity Rate is the most recent programmed value registered
in pacemaker memory. This value will be used even when it is not pertinent to the present
programmed mode.
D-6 Prodigy Reference Information