Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Dual Chamber Pacemaker Diagnostics
NOTE: Prodigy diagnostics are programmed using the diagnostic setups
listed below that set the diagnostic type and associated parameters listed
under “setup values”. Alternately, other programmable values can be
selected for the diagnostic type and parameters.
Diagnostic Setup Setup Values Other Programmable Values
A. Tachy Monitor
Type: High Rate Episode-A.
Method: Rolling Frozen
Detection Rate
1
: Upper Tracking Rate 80 - 400 ppm
or Upper Activity Rate,
whichever is greater
Detection
Number: 8 2 - 200
Termination
Number: 8 2 - 20
Collect EGM: ON OFF
Lead Monitor
2
: OFF ON,
ON + Polar. SW.
(A+V, A, or V)
V. Tachy Monitor
Type: High Rate Episode-V.
Method: Rolling Frozen
Detection Rate
1
: Upper Tracking Rate 80 - 400 ppm
or Upper Activity Rate,
whichever is greater
Detection
Number: 3 2 - 200
Termination
Number: 3 2 - 20
Collect EGM: ON OFF
Lead Monitor
2
: OFF ON,
ON + Polar. SW.
(A+V, A, or V)
D-4 Prodigy Reference Information
9879EN-D_9501043EN/CTC.qxd 5/10/99 11:39 Page D-4