Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Telemetry Functions
Listed below are Marker Channel telemetries, battery and lead telemetry,
diagnostics, and event summaries for each of the Prodigy pacemakers.
Marker Channel Telemetry
Prodigy Prodigy Prodigy Prodigy Prodigy
DR D VDD SR S
Marker Telemetry ✓✓✓✓✓
Sensor Telemetry ✓✓✓
Marker Channel ✓✓✓
Diagrams
EGM ✓✓✓✓✓
(A or V)
Battery and Lead Information
Prodigy DR, Prodigy D, Prodigy VDD, Prodigy SR, Prodigy S
Battery Status Value Tolerance*
Message “OK” or “Replace Pacer”
Voltage V ± 10%
Current µA ± 30%
Impedance Ohms ± 10% (± 100 ohms)
Est. Time to Replace Months
Atrial and/or Value Tolerance*
Ventricular Chambers
Pulse Width ms ± 1%
Amplitude V ± 10%
Output Energy µJ ± 45%
Lead Current mA ± 35%
Lead Impedance Ohms ± 20%
Pacing Polarity UNI/BI
* Based on a 200 to 1250 ohm load at the time of manufacture until ERI.
Magnet Mode Operations, Telemetry Functions, Diagnostics, and Event Summaries D-3
9879EN-D_9501043EN/CTC.qxd 5/10/99 11:39 Page D-3