Prodigy VDD Pacemakers’ Projected Longevity from Implant to ERI
Programmed Settings Percent Paced Projected Longevity (Years)
Prodigy VDD
Rate: 60 ppm (8168)
Mode: VDD
Pulse Width: 4.0 ms
Lead Impedance
600
ohms
Nominal Amplitude: 3.5 V 100% 9.8
50% 11.2
Low Amplitude: 2.5 V 100% 11.5
50% 12.4
Prodigy SR and Prodigy S Pacemakers’ Projected Longevity from Implant to ERI
Programmed Settings Percent Paced Projected Longevity (Years)
Prodigy SR Prodigy S
Rate: 60 ppm (8158, 8160/61/62) (8164/65/66)
Mode: VVIR VVI
Pulse Width: 0.4 ms
Lead Impedance
500 600 1000 500 600 1000
ohms ohms ohms ohms ohms ohms
Nominal Amplitude: 3.5 V 100% 7.8 8.2 9.3 7.9 8.4 9.4
50% 9.3 9.6 10.3 9.4 9.7 10.4
Low Amplitude: 2.5 V 100% 9.6 9.9 10.5 9.8 10.1 10.7
50% 10.5 10.7 11.1 10.7 10.9 11.3
WARNING: When the elective replacement indicator (ERI) is set, the pacemaker will continue to
operate at ERI conditions for a minimum of 3.0 months for the Prodigy VDD pacemaker and a
minimum of 2.5 months for the Prodigy SR and Prodigy S pacemakers until the battery depletes.
NOTE: The values above are based on calculations using deliverable battery capacity. These values
are estimates of longevity projections from implant to ERI and are intended to assist the clinician in
understanding the effects of various pacing conditions on battery longevity. These values should not
be interpreted as precise numbers.
Longevity Projections, Elective Replacement Indicator, and Battery Specifications C-3