Longevity Projections
Projected Longevity from Implant to ERI
The following tables show the estimated time from implant to the elective replacement indicator (ERI)
being set at nominal and low output settings for Prodigy pacemakers. The estimated time varies based
on the lead impedance, the programmed amplitudes, and the percent paced.
Prodigy DR and Prodigy D Pacemakers’ Projected Longevity from Implant to ERI
Programmed Settings Percent Paced Projected Longevity (Years)
Prodigy DR Prodigy D
Rate: 60 ppm (7860/61/62) (7864/65/66)
Mode: DDDR DDD
Pulse Width
Atrial: 0.4 ms Lead Impedance
Ventricular: 0.4 ms 500 600 1000 500 600 1000
ohms ohms ohms ohms ohms ohms
Nominal Amplitudes 100% 7.2 7.7 9.2 7.3 7.8 9.3
Atrial: 3.5 V
Ventricular: 3.5 V 50% 9.2 9.6 10.8 9.3 9.8 10.9
Low Amplitudes 100% 9.7 10.1 11.1 9.8 10.2 11.3
Atrial: 2.5 V
Ventricular: 2.5 V 50% 11.2 11.4 12.1 11.3 11.6 12.3
WARNING: When the elective replacement indicator (ERI) is set, the pacemaker will continue to
operate at ERI conditions for a minimum of 3 months until the battery depletes.
C-2 Prodigy Reference Information
9879EN-C_9501043EN/CTC.qxd 5/10/99 11:38 Page C-2