Specifications

7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
NOTE: Executing an Interrogate All command automatically saves the
present programmed parameter settings. From a parameter programming
screen, you can view the saved values applicable to that screen (select the
[
SHOW SAVED] button) or you can recall the values to the screen as pending
values for programming (select the [
SAVE
/
NOMINAL OPTIONS
...
] button,
then select the [
SAVED VALUES] button).
The INTERROGATE Function
From most function screens, you can interrogate the pacemaker for the data
associated with that screen. Even when you have used the Interrogate All
feature, it may be necessary to perform additional interrogations to update
the programmer as you conduct the patient session.
To execute the Interrogate command:
1. Position the programming head over the pacemaker.
2. Select the [
INTERROGATE] button. Continue to hold the programming
head steady until success of the interrogation is confirmed.
Interrogate Button
2-8 Using the Programmer
Ï