Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Shipping Parameter Settings
NOTE: Parameter settings not appearing for a specific Prodigy pacemaker are not applicable to that
device. Some parameter settings are dependent on mode or other parameters.
Parameter Prodigy DR Prodigy D Prodigy VDD Prodigy SR Prodigy S
(8158)
(7860/61/62) (7864/65/66) (8168) (8160/61/62) (8164/65/66)
Mode DDD DDD VDD VVI VVI
Mode Switch OFF
Lower (Pacing) 60 ppm 60 ppm 50 ppm 60 ppm 60 ppm
Rate
Upper Activity 120 ppm 120 ppm
Rate*
Upper Tracking 120 ppm 120 ppm 120 ppm
Rate
Paced AV (PAV) 150 ms 150 ms
Sensed AV (SAV) 120 ms 120 ms 120 ms
RAAV OFF OFF OFF
PVARP 310 ms 310 ms 310 ms
Ventricular 230 ms 230 ms 230 ms 330 ms 330 ms
Refractory
Period
PMT Intervention OFF OFF OFF
Atrial
Amplitude 3.5 V 3.5 V
Pulse Width 0.4 ms 0.4 ms
Sensitivity 0.5 mV 0.5 mV 0.25 mV
Blanking 220 ms 220 ms 220 ms
(PVAB)
Pacing Polarity See package See package
label label
Sensing Polarity See package See package Bipolar
label label
B-2 Prodigy Reference Information
9879EN-B_9501043EN/CTC.qxd 5/10/99 11:37 Page B-2