Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Basic Description
The Medtronic Prodigy family of pacemakers consists of multi-
programmable, implantable pacemakers. The five pacemakers in the
Prodigy family offer a wide range of therapeutic features and diagnostic
capabilities based on a common platform. Physicians may choose the
appropriate pacemaker for the patient and expect similar programming and
follow-up operations between the five pacemakers.
Medtronic programmers allow the clinician to change pacing therapies and
assess pacing and diagnostic data by communicating with the pacemaker
via telemetry. In addition, the clinician can perform system functions such
as interrogating the pacemaker, measuring stimulation thresholds, and
inhibiting the pacemaker.
Prodigy DR pacemakers are dual chamber, activity-based rate responsive
DDDR pacemakers.
■
Model 7860 is compatible with IS-1* bipolar/unipolar leads.
■
Model 7861 is compatible with 5 or 6 mm leads.
■
Model 7862 is compatible with low-profile 3.2 mm bipolar leads and
IS-1* bipolar/unipolar leads.
Prodigy D pacemakers are dual chamber, DDD pacemakers.
■
Model 7864 is compatible with IS-1* bipolar/unipolar leads.
■
Model 7865 is compatible with 5 or 6 mm unipolar leads.
■
Model 7866 is compatible with low-profile 3.2 mm bipolar leads and
IS-1* bipolar/unipolar leads.
Prodigy VDD pacemakers are dual chamber VDD pacemakers that utilize
a single lead pacing system. In addition, single chamber, activity-based rate
responsive VVIR pacing is ON/OFF programmable. The Prodigy VDD
pacemakers are designed for use with the CapSure VDD (Model 5032)
lead, which provides bipolar atrial sensing and ventricular pacing and
sensing. The pacemaker’s ventricular port is IS-1* compatible. The atrial
port is a bipolar connector which meets the mechanical requirements of the
IS-1 standard but is only designed to sense.
* IS-1 refers to an International Connector Standard (See Document No. ISO 5841-3; 1992)
whereby pulse generators and leads so designated are assured of a basic mechanical fit.