9760 Programmer
If you are using a 9760 Programmer, the default setting for the Assist
function is Normal. That is, the Assist function will not be in effect when
you start a patient session. If you have trouble receiving telemetry or ob-
taining a programming confirmation, enabling the Assist function can
facilitate locating a proper programming head position. As stated above,
this function results in the appearance of extraneous artifacts on the pa-
tient’s ECG.
Procedure for Selecting the Position Head Assist Setting
1. Select [
SPECIAL
] from the menu buttons at the top of the screen.
2. Select
POSITION HEAD ASSIST
from the options displayed.
3. From the window showing two buttons:
Select [
ENHANCED
] to enable the Assist function.
Or select [
NORMAL
] to disable the Assist function.
Position Head Assist
Option