Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
How VSP Operates
When VSP is programmed ON, a ventricular sense within 110 ms after an
atrial pacing pulse results in a ventricular pacing pulse at 110 ms (see
illustration below) or at the programmed PAV interval, whichever expires
first. If the sensed event was, in fact, a ventricular depolarization, the
ventricular pacing pulse at 110 ms falls harmlessly into the absolute
refractory period of the ventricle.
Ventricular Safety Pacing (VSP)
14-10 Understanding Pacemaker Operations
Act.
P
P
Safe
DDDR
A
V
S
110
ms
Act.
P
P
Act.
P
P