Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Using the Command Buttons
Listed below is a description of each of the command buttons located at the
bottom of the display screen. Refer to the listed chapter for the procedure
associated with using a particular command button.
Programming Emergency Parameters — See Chapter 2.
Selecting the Emergency command overrides any function and immediately
programs the pacemaker to a fixed set of permanent values that provide
high output pacing in the VVI mode.
Canceling the Programming Head Magnet — See Chapter 2.
Selecting the Cancel Magnet command cancels the effect of the program-
ming head magnet and allows the pacemaker to operate in its permanently
programmed state while the programming head is in position. For example,
operation of a pacemaker permanently programmed to the DDD mode will
change from the DOO magnet mode to the DDD mode when this command
is executed.
Interrogating the Pacemaker — See Chapter 2.
Selecting the Interrogate command from a particular function screen
retrieves from the pacemaker the data related to that function.
Interrogating All Data — See Chapter 2.
The Interrogate All command retrieves all data (parameter, battery/lead, and
diagnostic) from the pacemaker. This button appears only on the Pacing
Parameters screen.
Returning to the Pacing Parameters Screen
On screens other than the Pacing Parameters screen, this button appears in
place of [
INTERROGATE ALL
]. Selecting this button returns you directly to
the Pacing Parameters screen.
Programming Command — (Covered in the respective procedures).
This command executes a temporary or permanent programming transmis-
sion based on the procedure being conducted. This button appears only
when programming is allowed.
EMERGENCY
CANCEL
MAGNET
INTERROGATE
INTERROGATE
ALL
PACING
PARAMETERS
PROGRAM
1-10 Using the Programmer
(not pictured)