The example below shows how the Lead Trend report is presented on the
programmer. The report can be printed out or displayed on the
programmer.
Example of Lead Trend Diagnostic
Overriding Other Diagnostics
If the Lead Trend is activated because the Lead Monitor detects out-of-
range lead impedance(s), any programmed diagnostic is disabled, and its
data are overwritten by Lead Trend data.
For Further Information
Refer to Chapter 8 “Collecting and Retrieving Diagnostic Data” for
information on configuring the Lead Trend diagnostic and collecting its
data.