Specifications

Sensed AV Interval (SAV). SAV follows an atrial sensed event in atrial
synchronous pacing modes (DDDR, DDD, and VDD). The SAV
interval duration may differ from the programmed value due to:
- Rate Adaptive AV operation or
- Wenckebach operation, where the SAV is extended to avoid
violation of the Upper Tracking Rate or the total atrial refractory
period while tracking a fast intrinsic atrial rate (refer to “High Rate
Atrial Tracking” later in this chapter).
Example of SAV Interval Operation
Pacemaker Timing 12-9
9879EN-12_9501043EN/CTC.qxd 5/10/99 11:27 Page 12-9
SAV Interval SAV Interval
S
PP
DDD
A
V
S
SAV SAV
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A