Specifications

7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
To select an option:
Without touching the screen, move the tip of the selector pen to a
position directly over any part of the boldface option label (may be a
word or number); then press the pen against the screen.
Audible Signals
Certain events in the operation of the programmer will result in an audible
signal of one or two tones, or “beeps.” These signals are intended to alert
the user to the success or failure of an action.
Single Beep = Success
A single beep following an Interrogate command indicates that the
interrogation was successful.
A single beep following a Program command indicates that the pro-
gramming was confirmed.
Two Beeps = Failure
Two beeps following an Interrogate command indicate that the inter-
rogation was not successful.
Two beeps following a Program command indicate that the program-
ming was not confirmed.
Two Low-Tone Beeps
Two low-tone beeps sound when the selected command cannot be
executed.
Single Short Beep
A short beep that coincides with pressing the
INTERROGATE
or
PROGRAM
key on the programming head indicates that the programmer acknowledges
the keystroke. If the keystroke is inappropriate, two low-tone beeps will
sound.
Using the Programmer with Prodigy™ Pacemakers 1-7