Specifications

7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Rates
Overview
The following programmable rates control timing in Prodigy pacemakers:
Normal operating rates:
- Lower Rate
- Upper Tracking Rate
- Upper Activity Rate
Other operating rates:
- Hysteresis Rate (single chamber demand modes)
Additionally, rates calculated by the pacemaker are used for some
operations. These are:
sensor-indicated rate
mean atrial rate
The other operating rates are described in Chapter 14, along with the
functions that use them. The normal rates are described in this chapter.
A-A Timing
In all modes that pace the atrium, Prodigy pacemakers time from atrial
event to atrial event when possible (A-A timing). This timing method
mimics a natural sinus rhythm, producing A-A intervals that are nearly
equal, except when timing is interrupted by:
PACs in the DDIR and DDI modes,
PVCs in the DDDR, DDD, DDIR, and DDI modes, or
Ventricular sensed events during the VA interval in the DVIR and DVI
modes.
VA intervals vary due to adjustments by A-A timing operations designed to
achieve sensor-indicated or lower rate operation in the presence of varying
AV conduction.
12-2 Understanding Pacemaker Operations