Specifications

DOOR, AOOR, and VOOR Mode Operation
The DOOR, AOOR, and VOOR modes operate as follows:
The DOOR mode provides asynchronous AV sequential pacing at the
sensor-indicated rate. Intrinsic events are ignored.
The AOOR and VOOR modes provide single chamber pacing at the
sensor-indicated rate. Intrinsic events are ignored.
Example of DOOR Mode Operation
Pacing Modes 10-33
V
P
Sensor-indicated
Interval
Parameters:
Lower Rate = 60 ppm (1000 ms)
Sensor-indicated Rate = 90 ppm (667 ms)
PAV Interval = 200 ms
200 ms
A
P
V
P
Sensor-indicated
Interval
A
P
V
P
Sensor-indicated
Interval
A
P
A
P
V
P
Sensor-indicated
Interval
A
P