Specifications
Example of DDIR Mode Operation.
Pacing Modes 10-13
V
P
A
P
A
S
A
P
A
P
Sensor-indicated
Interval
Parameters:
Lower Rate = 60 ppm (1000 ms)
Sensor-indicated Rate = 90 ppm (667 ms)
PAV Interval = 200 ms
Sensor-indicated
Interval
V
P
V
P
Sensor-indicated
VA Interval
200 ms
V
P
A
P
A
P
Sensor-indicated
Interval
9879EN-10_9501042_CTC.qxd 12/13/01 3:57 PM Page 10-13
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.