c) Select the pacemaker type, [
SINGLE CHAMBER MODELS
] or [
DUAL
CHAMBER MODELS
].
d) Select the desired Prodigy pacemaker group from those listed.
Selecting the pacemaker group results in a data loading process
that will take several seconds. The Site Nominal Setup screen for
Pacing Parameters is then displayed (see example below).
If Site Nominal values have not been set up for the selected models, the
screen will show Medtronic nominal values with all values pending.
The screen otherwise shows the present Site Nominal values.
3. To establish the desired set of Site Nominal values, make the required
changes to the parameters settings displayed:
a) Select a parameter you want to change.
For pacing parameters, select Pacing Mode first if you intend to
change it.
b) From the window of value options, select the desired value.
NOTE: Selecting a value of
ON
for Mode Switch will display the
present setting for Detect Rate within the Mode Switch value
window. To choose a different setting, select Detect Rate to open
its value window. After choosing the desired rate value, select
[
ACCEPT VALUE] to close the Mode Switch window (see page 6-5).
Site Nominal Setup
Screen (example)
Programmer Setup Options 9-7