EGM/ECG Signal Calibration
Depending on the displayed waveform trace, selecting the calibrate option
also generates a voltage reference diagram that applies as follows:
If you are using a 9760 programmer, the voltage reference diagram
appears only in the telemetered EGM trace displayed or printed by an
externally connected recorder or monitor.
If you are using a 9790 programmer, the voltage reference diagram
appears in both the ECG and telemetered EGM traces displayed and
printed by the programmer as well as in the ECG and EGM traces
displayed or printed by the externally connected recorder or monitor.
The signal calibration diagram provides a voltage reference for determining
waveform amplitudes.
5 mV Æ
1 mV Æ
0 mV Æ
9760 Programmer 9790 Programmer