Specifications
7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Clearing Recorded Data From the Pacemaker
The Clear Data command lets you clear the Diagnostic data presently
accumulated in pacemaker memory so that new data collection can begin.
The Clear Data command does not alter the present Diagnostic
parameter settings.
New data collection begins 30 seconds after you lift the programming
head from over the patient’s pacemaker.
For a permanent record of the present data, print the data as described
in the previous section before you execute the Clear Data command.
To clear the Diagnostic data accumulated in pacemaker memory:
1. Select the [
CLEAR DATA
] button, which appears on the right side of the
Diagnostic Parameters screen (see page 8-13) and in the same location
on the Diagnostics graph display screen.
2. Position the programming head over the patient’s pacemaker and select
the [
CLEAR DATA
] button from the window now showing on the screen.
NOTE: Any one of the following will also clear the recorded Diagnostic
data:
Programming any Diagnostic parameter.
Conducting an Activity Threshold test or Exercise test.
The occurrence of a pacemaker electrical reset.
8-20 Using the Programmer