Specifications

7 x 9 inches (178 mm x 229 mm)
Medtronic Confidential
Refer to the “Reference man.” category in doc#163256 for Printing
Instructions.
196407001 Rev A
Data collection begins 30 seconds after you lift the programming head.
(Data collection is suspended while the programming head is in position
over the patient’s pacemaker.)
Procedure for Manually Selecting the Parameter Settings
As an alternative to using the Setup feature, you can manually select each
of the parameter settings required to program the Diagnostics function.
NOTE: You can start this procedure
by selecting an appropriate Setup option
as described above and then using this
procedure to modify the preselected
pending parameter values.
1. From the Diagnostic Parameters
screen, select
Type
from the
parameter list.
2. From the window of Type options,
select the option for the type of
data you want to collect. (If you are
programming the Diagnostics
function off, select
Off
and skip to
step 4.)
3. The screen now displays the
parameters pertinent to the selected
Type setting.
Select a desired setting for each of
the listed parameters by selecting
the parameter and then selecting
the desired setting from the value
window.
The shaded rectangles, which may
appear depending on the previous
programmed status, indicate where
values must be selected before
programming will be allowed.
4. To program the selected pending
settings, position the programming
head and select [
PROGRAM
].
Data collection begins 30 seconds after you lift the programming head.
(Data collection is suspended while the programming head is in position
over the patient’s pacemaker.)
8-16 Using the Programmer