NOTE: If low battery voltage has set the pacemaker ERI (elective replace-
ment indicator), the programmer will not allow you to program the
Diagnostics function.
Diagnostic Parameter Screen Buttons
The buttons located at the right of the parameter display may be useful
during the programming procedure described on the next page.
When you select new parameter values to program, these
“pending” values are displayed in place of the interrogated
present settings. To temporarily recall the present settings,
press and hold this button.
If you have saved a set of parameter values (see below), this
button will appear on the screen. To temporarily display these
saved values, press and hold this button.
If the Diagnostics function was previously programmed to
collect data and if this data has been interrogated, selecting
this button will display a graph of the data retrieved from
pacemaker memory.
Selecting this button displays an option to clear the Diagnos-
tics data presently accumulated in pacemaker memory (see
page 8-20).
Selecting this button displays options for: 1) clearing all
pending values from the parameter screen, 2) saving a set of
parameter values, and 3) recalling the set of saved values or
nominal values (Medtronic or Site) to the screen as pending
values. (Refer to “Programming Nominal and Saved Values”
on page 6-6 for instructions on using these options.)
SHOW
PRESENT
SHOW
SAVED
SHOW GRAPH
CLEAR DATA
SAVE/
NOMINAL
OPTIONS...